|
Product Details:
|
Assay Time: | 2 Hours | Detection Range: | 0.1 - 10 Ng/mL |
---|---|---|---|
Expiration Date: | 18 Months | Manufacturer: | Biovantion Inc |
Method: | Sandwich ELISA Kit | Organism: | Mouse |
Package Size: | 96 Tests/Kit | Precision: | Varies Depending On Target Analyte |
Sample Type: | Biological Samples | Size: | 96 Tests |
Specificity: | Highly Specific | Storage Conditions: | Store At Recommended Temperature |
Test Format: | ELISA | Test Sample Type: | Serum, Plasma |
Usage: | For Research Use Only | ||
Highlight: | Research Use Only Test Kit,Ruo Diagnostic Kit,Ruo Assay Kit |
RUO Antibodies to Herpes Simplex Virus (HSV Ⅰ) IgM ELISA Test Kit (Serum/Plasma)
This kit is a qualitative detection of human serum/plasma of HSVⅠ IgM . The kit is suitable for clinical screening and diagnosis of HSVⅠ infection in serum/plasma.
MAIN COMPONENTS
Materials provided with the kit:
Components | 96T/Box | 480T/Box | ||
Coated Microtiter Plate | 1 bag | 8*12 | 5 bags | 8*12 |
Conjugate | 1 vial | 6.5 mL | 5 vials | 6.5 mL |
Wash Buffer (40X) | 1 vial | 20 mL | 5 vials | 20 mL |
Sample Diluent | 1 vial | 11 mL | 5 vials | 11 mL |
Substrate A | 1 vial | 7 mL | 5 vials | 7 mL |
Substrate B | 1 vial | 7 mL | 5 vials | 7 mL |
Stop Solution | 1 vial | 6 mL | 5 vials | 6 mL |
Negative Control | 1 vial | 1 mL | 5 vials | 1 mL |
Positive Control | 1 vial | 1 mL | 5 vials | 1 mL |
Closure Plate Membrane | 3 sheets | 15 sheets |
Note: different batches of reagent kit, and different component cannot be exchanged for use. Once open, stable for 3 months at 2-8°C.
TEST PROCEDURE
1. All reagents should be allowed to reach room temperature for 15 minutes before use.
2. Dilute the wash buffer at the rate of 1:40 dilution with distilled water before use.
3. Add 100μL Sample Diluent in the corresponding well (Do not add in the blank well, negative control wells and positive control well.) The sample should be corresponding to the number of micro plate, each plate should be provided with negative control 2 wells, positive control 1 well and blank control 1 well. (If detect with dual wavelength detection, setting no blank control well is allowed).
Note: Use a separate disposal pipette tip for each specimen, Negative and Positive Control to avoid cross contamination.
4. Add 10μL sample in the corresponding well, mix thoroughly by using the pipette, add 100μL negative control and positive control to negative control wells and positive control well (Do not add in the blank well).
5. Shake gently to mix for 30s. Incubate at 37°C for 20 minutes with the sealing plate membrane sealing the plate.
6. At the end of the incubation, remove and discard the plate cover. Take out, add wash buffer to each well for 20 seconds. Repeat 5 times. After the final washing cycle, turn the plate over onto blotting paper or clean towel, and tap it to remove any remainders.
7. Respectively adding Conjugate 50μL (Do not add in the blank well) Incubate at 37°C for 20 minutes with the sealing plate membrane sealing the plate.
8. Repeat the wash step for 5 times as in step 6.
9. Add Substrate A (50µL) and Substrate B (50µL) (Do not add in the blank well). Incubate at 37°C for 10 minutes with the sealing plate membrane sealing the plate.
10. Add 50μL Stop Solution to each well (Do not add in the blank well). Mix gently by shaking, read the absorbance within 10 minutes after stopping the reaction. Calibrate the plate reader with the Blank well and read the absorbance at 450nm. If a dual filter instrument is used, set the reference wavelength at 630nm. Set no blank wells is allowed if use dual wavelength to detect. Calculate the Cut-off value and evaluate the results.
Contact Person: Mr. Steven
Tel: +8618600464506